Plasma volume measurement in septic patients using an albumin dilution technique: comparison with the standard radio-labelled albumin method

  • Experimental
  • Published:
Intensive Care Medicine Aims and scope Submit manuscript

Abstract

Objective

To investigate a technique using 20% albumin for measurement of plasma volume in critically ill patients.

Design and setting

Laboratory and clinical investigation in the adult intensive care unit and anaesthetic laboratories of a university hospital.

Patients

12 patients fulfilling ACCP/SCCM criteria for septic shock.

Interventions and measurements

Each patient received 125I-labelled albumin, and the volume of distribution was measured at 1 and 10 min. The accepted standard plasma volume measurement (98% of the 10-min volume of distribution) was calculated. Immediately thereafter 200 ml 20% human albumin was given. Albumin concentrations were measured before and 1 min after this 40-g bolus, and the volume of distribution calculated using a formula that corrected for the 200 ml fluid in which the albumin was dissolved.

Results

Plasma volumes measured using the albumin dilution technique at 1 min were smaller than the standard technique by 110±280 ml; limits of agreement were from −660 to +440 ml (−16% to +11%). Plasma volumes measured by 125I-albumin at 1 min were smaller than the standard by 120±110 ml; limits of agreement were from −330 to +100 ml (−8 to +2%).

Conclusions

Non-labelled albumin can be used easily and quickly to measure a plasma volume in ICU patients and gives a moderately accurate estimate when compared with the 125I-labelled albumin methods. The normal isotope method over-estimates plasma volumes in septic patients because excessive transcapillary escape of albumin is inadequately compensated for by the standard correction factor.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Bone RC, Balk RA, Cerra FB, Dellinger RP, Feim AM, Knaus WA, Schein RM, Sibbald WJ (1992) Definitions for sepsis and organ failure and guidelines for the use of innovative therapies in sepsis. The ACCP/SCCM Consensus Conference Committee. American College of Chest Physicians/Society of Critical Care Medicine. Chest 101:1644–1645

    CAS  PubMed  Google Scholar 

  2. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I:307–310

    Google Scholar 

  3. Tchaikowsky K, Meisner M, Durst R, Rugheimer R (1997) Blood volume determination using hydroxyethyl starch: a rapid and simple intravenous injection method. Crit Care Med 25:599–606

    Article  PubMed  Google Scholar 

  4. Van Kreel BK, van Beek E, Spaanderman ME, Peeters LL (1998) A new method for plasma volume measurements with unlabelled dextran-70 instead of 125-I-labelled albumin as an indicator. Clin Chim Acta 275:71–80

    Article  PubMed  Google Scholar 

  5. Brown MA, Mitar DA, Whitworth JA (1992) Measurement of plasma volume in pregnancy. Clin Sci (Colch) 83:29–34

    Google Scholar 

  6. Anthony MY, Goodall SR, Papouli M, Levene MI (1992) Measurement of plasma volume in neonates. Arch Dis Child 67:36–40

    CAS  PubMed  Google Scholar 

  7. Gillen CM, Takamata A, Mack GW, Nadel ER (1994) Measurement of plasma volume in rats with use of fluorescent-labelled albumin molecules. J Appl Physiol 76:485–489

    CAS  PubMed  Google Scholar 

  8. Haruna M, Kumon K, Yahagi N, Watanabe Y, Ishida Y, Kobayashi N, Aoyagi T (1998) Blood volume measurement at the bedside using ICG pulse spectrophotometry. Anesthesiology 89:1310–1312

    Article  CAS  PubMed  Google Scholar 

  9. Thomsen JK, Fogh-Andersen N, Bülow K, Devantier A (1991) Blood and plasma volumes determined by carbon monoxide gas,99mTc-labelled erythrocytes, 125I-albumin and the T 1824 technique. Scand J Clin Invest 51:185–189

    CAS  PubMed  Google Scholar 

  10. Farjanel J, Denis C, Chatard JC, Geyssant A (1997) An accurate method of plasma volume measurement by direct analysis of Evans blue spectra in plasma without dye extraction: origins of albumin-space variations during maximal exercise. Eur J Appl Physiol 75:75–82

    Article  CAS  Google Scholar 

  11. Cochrane Injuries Group (1998) Human albumin administration in critically ill patients: systematic review of randomised controlled trials. BMJ 317:235–240

    PubMed  Google Scholar 

  12. Bunn F, Alderson P, Hawkins V (2000) Colloid fluids for fluid resuscitation. Cochrane database systematic review reference 2. pages CD001319

Download references

Author information

Authors and Affiliations

Authors

Corresponding authors

Correspondence to Michael P. Margarson or Neil C. Soni.

Additional information

Dr. Margarson’s research post (SIMS Fellow, Magill Department of Anaesthesia) was funded by Smiths Industries Medical Systems

Electronic Supplementary Material

Rights and permissions

Reprints and permissions

About this article

Cite this article

Margarson, M.P., Soni, N.C. Plasma volume measurement in septic patients using an albumin dilution technique: comparison with the standard radio-labelled albumin method. Intensive Care Med 31, 289–295 (2005). https://doi.org/10.1007/s00134-004-2481-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00134-004-2481-4

Keywords

Navigation