Original Article
Effective Dose of Dual-Energy X-Ray Absorptiometry Scans in Children as a Function of Age

https://doi.org/10.1385/JCD:8:4:415Get rights and content

Abstract

Effective dose, a parameter utilized to assess biological risk related to radiation exposure, may be used to evaluate risk associated with dual-energy X-ray absorptiometry (DXA). We estimated the effective dose from DXA (Hologic QDR 4500A) scans of the lumbar spine (fast array mode), total body, hip (fast array mode), and forearm for children ages 1, 5, 10, and 15 yr and for adults. Entrance dose incorporating backscatter was determined for each scan type. Depth-dose curves were derived using Plexiglas slabs simulating tissue attenuation. Organ depth was estimated using pediatric phantom models. For all scan types, the effective dose decreased as age increased. The effective dose values for a 1-yr-old and an adult, respectively, were 4.7 μSv and 2.2 μSv for a lumbar spine scan performed in fast array mode, 3.4/3.5 μSv and 1.8/2.1 μSv (male/female) for a total body scan, and 0.14 μSv and 0.03 μSv for a forearm scan. There were marked sex differences in the effective dose associated with hip scans (fast array mode) ranging from 15.2 μSv for a 1-yr-old male to 4.6 μSv for an adult female. A comprehensive uncertainty analysis indicated that the effective dose values were reliable within a factor of 3. With the exception of the hip scans in 1- and 5-yr-olds, the effective doses were below the negligible individual dose limit of 10 μSv/yr.

References (12)

  • CF Njeh et al.

    Radiation exposure in bone mineral density assessment

    Appl Radiat Isot

    (1999)
  • International Commission on Radiological Protection

    1990 Recommendations of the International Commission on Radiological Protection

    Ann ICRP

    (1991)
  • National Council on Radiation Protection and Measurements

    Limitation of Exposure to Ionizing Radiation. No. 116

    (1993)
  • MK Lewis et al.

    Patient dose in dual X-ray absorptiometry

    Osteoporos Int

    (1994)
  • CF Njeh et al.

    Radiological assessment of a new bone densitometer—the Lunar EXPERT

    Br J Radiol

    (1996)
  • E Bezakova et al.

    Absorbed dose measurements in dual energy X-ray absorptiometry (DXA)

    Br J Radiol

    (1997)
There are more references available in the full text version of this article.

Cited by (51)

  • Update on bone density measurements and their interpretation in children and adolescents

    2018, Best Practice and Research: Clinical Endocrinology and Metabolism
  • Energy balance and fitness in adult survivors of childhood acute lymphoblastic leukemia

    2015, Blood
    Citation Excerpt :

    Waist circumference in centimeters was measured with a Gulick tape measure and divided by height to characterize waist-to-height ratio. Fat and lean mass were measured with dual x-ray absorptiometry (QDR 4500, software version 13.3:3; Hologic, Bedford, MA) in the total-body scanning mode.13-17 The scanner was calibrated weekly with known phantoms to minimize machine drift.

  • Vitamin D status is associated with bone mineral density and bone mineral content in preschool-aged children

    2015, Journal of Clinical Densitometry
    Citation Excerpt :

    The cDXA spine phantom with known density (1.024 g/cm2) yielded a BMD CV of 0.3% (244 scans). Effective X-ray dose for all scans combined was 8.6 μSV (23). Of the 78 children, all were successfully scanned with no movement breaks in the images for LS 1–4, 71 for whole body and 59 for forearm.

  • Dose absorption in lumbar and femoral dual energy x-ray absorptiometry examinations using three different scan modalities: An anthropomorphic phantom study

    2013, Journal of Clinical Densitometry
    Citation Excerpt :

    Dual energy X-ray absorptiometry (DXA) is currently the standard technique to measure bone mineral density and to diagnose osteoporosis, according to the World Health Organization guidelines (1,2).

View all citing articles on Scopus
View full text